With the SleepOne Bilevel Auto,patients could also be treated with auto adjusting bilevel modes.Adjustable triggering sensitivity increases patient compliance,and makes the device useful for different patient groups. Inspiration Trigger is adjustable from 2 l/min to 20 l/min flow. To increase the device-patient synchronization, both IPAP-EPAP and EPAP-IPAP pressure change curves could be rounded in 5 steps. This setting improves the patient comfort thus it increases the synchronization.
Mode 1 – Auto Bilevel Constant Delta PThis Mode starts with the minimum EPAP, Maximum IPAP, Pressure Support, Ti Min and Ti Max set by the physician. Device auto titrates the IPAP pressure and defines the EPAP according to fixed parameter Pressure Support(IPAP-EPAP). So there is always a constant pressure difference between IPAP and EPAP. Device responds to Apnea, Snore, Hypopnea, RERA(Respiratory Effort Related Arousal)
Mode 2 – Auto Bilevel Variable Delta PThis Mode starts with the minimum EPAP,Maximum IPAP,Pressure Support,Min Pressure Support,Max Pressure Support,Ti Min and Ti Max set by the physician.Device auto titrates both IPAP and EPAP pressures.So there is a changing Pressure Support.Device responds to Apnea,Snore,Hypopnea,Flow Limitation(RERA).
Available Modes:CPAP,Auto Bilevel Constant,Auto Bilevel Variable
Inspiration Trigger:2 l/min–20 l/min adjustable
Expiration Trigger:%10–%60 of inspired flow
EPAP Rounding:0–5*
Report Software:SleepONE Software
* Caution: At EPAP Rounding setting 0,pressure drops to 0 cmH2O for few miliseconds.
Mode 1 – Auto Bilevel Constant Delta PThis Mode starts with the minimum EPAP, Maximum IPAP, Pressure Support, Ti Min and Ti Max set by the physician. Device auto titrates the IPAP pressure and defines the EPAP according to fixed parameter Pressure Support(IPAP-EPAP). So there is always a constant pressure difference between IPAP and EPAP. Device responds to Apnea, Snore, Hypopnea, RERA(Respiratory Effort Related Arousal)
Mode 2 – Auto Bilevel Variable Delta PThis Mode starts with the minimum EPAP,Maximum IPAP,Pressure Support,Min Pressure Support,Max Pressure Support,Ti Min and Ti Max set by the physician.Device auto titrates both IPAP and EPAP pressures.So there is a changing Pressure Support.Device responds to Apnea,Snore,Hypopnea,Flow Limitation(RERA).
Available Modes:CPAP,Auto Bilevel Constant,Auto Bilevel Variable
Inspiration Trigger:2 l/min–20 l/min adjustable
Expiration Trigger:%10–%60 of inspired flow
EPAP Rounding:0–5*
Report Software:SleepONE Software
* Caution: At EPAP Rounding setting 0,pressure drops to 0 cmH2O for few miliseconds.